Z-0309-2021: Critical Care Decontamination System (CCDS) Compatible N95 respirators - Product Usage: Used in deconta

Recall record Z-0309-2021 for Battelle Memorial Institute. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0309-2021
Company in FDA record
Battelle Memorial Institute
FDA record creation date
Oct 29, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Masks processed at one site were not maintained at levels of condensation during a portion of the decontamination cycle.

Source facts

Action
Customers were notified via email on 8/13/20: Dear CCDSTM Customer, Recently, we became aware that masks processed at the Providence, RI CCDSTM site were not maintained at levels of condensation during a portion of the decontamination cycle. The first phase (gassing phase) operated as intended. The next phase (dwell phase) had a gradual decline of condensation. The CCDSTM team has corrected this and has confirmed that the system is operating correctly. The normal CCDSTM decontamination cycles operate above the levels of regulatory guidelines. To ensure that our process decontaminated the masks effectively, our research team replicated this process cycle, including the deviated dwell phase, with a similar number of masks in the chamber. As expected, the results showed a 6-log reduction. Though this was a deviation to an internal operating procedure, out of caution, we are asking you to return your masks to the CCDSTM site to be reprocessed. We will send you replacement masks if you need them while your masks are being decontaminated. Battelle CCDSTM is committed to maintaining the highest standards for our customers. Please contact us with any questions. Sincerely, Battelle CCDSTM Team Call us at 614-424-4373 ccdscustomerservice@battelle.org
Initiated
2020-08-13
Root cause
Under Investigation by firm
Terminated
2021-02-03
Distribution
US Nationwide distribution including in the states of CT, MA, RI.
Recall number
Z-0309-2021
Recall status
Terminated
CDRH record ID
183449
FDA record created
2020-10-29
Recall event number
86347

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗