Open original FDA record ↗Recall record
- FDA identifier
- Z-0310-2012
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA classification date
- Nov 30, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The cobas c 311 analyzer software may incorrectly calculate the volume of the ISE Reference Electrolyte solution remaining on the analyzer. In some cases the incorrect results obtained will not be flagged with any data alarms.
Source facts
- Action
- October 26, 2011 Urgent Medical Device Corrections were sent to all direct accounts in the USA that have installed the cobas c 311 analyzer and included a faxback form to be completed and faxed to 1-877-766-7452. Consignees were instructed to temporarily replace the ISE Reference Electrolyte bottle based upon the frequency calculated from the maximum number of ISE samples run per day and activate the yellow threshold alarm on the system. Customers with question and concerns are directed to call Roche Diagnostics Technical Support at 1-800-428-2366, available 24-hours, 7 days a week.
- Initiated
- 2011-10-26
- Root cause
- Software design
- Terminated
- 2012-07-09
- Distribution
- Worldwide Distribution -- USA including the states of Maryland and Texas.
- Recall number
- Z-0310-2012
- Recall status
- Terminated
- CDRH record ID
- 104939
- FDA classified
- 2011-11-30
- Recall event number
- 60282