Skip to content

Public FDA record

Z-0310-2021: A Li-Gator Single Use Laparoscopic Clip Applier, 10 mm X 29 cm - Product Usage: intended for use on tubul

Recall record Z-0310-2021 for Genicon, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0310-2021
Company in FDA record
Genicon, Inc.
FDA record creation date
Oct 30, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Device malfunction causing the applier to jam, and clip loading failures.

Source facts

Action
On 9/17/20 the firm sent a notification to its consignees with the following instructions: Your assistance is appreciated and necessary to prevent any potential harm from the use of affected product. This recall is being carried out to the user level. If you have further distributed this product, please identify your customers and notify them at once of this product recall. Your notification to your customers may be enhanced by including a copy of this recall notification. Please complete and also encourage your customers to complete and return the enclosed response form as soon as possible. Return the form by e-mail to customerservice@geniconendo.com. In addition, if you or your customers have any of the affected devices listed in this letter, and the product is located within the United States of America, please contact Genicon (contact information listed below) to coordinate the return of the product associated with this action at no additional cost.
Initiated
2020-09-17
Root cause
Process design
Distribution
Worldwide distribution - US Nationwide distribution including in the states of FL, TN, MO, CA, GA, and IN. The countries of Chile, Croatia, Dubai (UAE), Equador, Egypt, France, Germany, Greece, Guatemala, Hong Kong, Iraq, Ireland, Israel, Jordan, Kuwait, Lebanon, Malta, Mexico, Morocco, Pakistan, Palestinian, Quatar, Saudi Arabia, Slovenia, South Africa, Switzerland, Tunisia, United Kingdom, and Venezuela.
Recall number
Z-0310-2021
Recall status
Open, Classified
CDRH record ID
183658
FDA record created
2020-10-30
Recall event number
86412

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗