Open original FDA record ↗Recall record
- FDA identifier
- Z-0310-2022
- Company in FDA record
- DRG International, Inc.
- FDA record creation date
- Nov 29, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Firm has identified two lots of product that do not meet product shelf life, resulting in failed Controls. The clinical impact may be a delay in diagnosis and a consumption of patient samples.
Source facts
- Action
- On 10/18/2021, DRG issued Medical Device Recall Notification letters and response forms to the End-user through confirmation of the Distributor. All customers were notified via e-mail. Actions Required: 1) Please immediately discontinue use or distribution of the affected lots. 2) If you have further distributed the affected product lots, please identify and immediately notify those customers at once of this product recall by providing a copy of the recall notice and response form. 3) If you have relabeled or rebranded the affected product lots, please follow your internal recall process and notify the appropriate regulatory bodies as required. 4) Disposal of all affected lots should be in accordance with applicable national, st ate, and local laws and regulations. 5) Complete and return the response form.
- Initiated
- 2021-10-18
- Root cause
- Nonconforming Material/Component
- Distribution
- Domestic: NY; International: TZ, VN, BD, CO, TT, YE, BO, DE, SV, GE, PY, TG, RU/UA, NI, DO, NG, EG & CM
- Recall number
- Z-0310-2022
- Recall status
- Open, Classified
- CDRH record ID
- 190088
- FDA record created
- 2021-11-29
- Recall event number
- 88949