Open original FDA record ↗Recall record
- FDA identifier
- Z-0311-2012
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA classification date
- Nov 30, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Internal investigations have revealed a potential printer malfunction on the Cobas c 111 analyzer, where one or more numbers or characters can be missing on the hard copy result printout, or the printer may skip leaving blanks which may result in an incorrect print layout. The missing number of characters can be in any position, therefore it could be in the middle of a result, (e.g. 48.53 mmol in
Source facts
- Action
- Roche Diagnostics Operations, Inc sent a Urgent Medical Device Correction dated October 18, 2011, via UPS Ground (receipt required) to all affected customers and one distribution center, with instructions to disable the Auto Print Result feature until new software is available. All printed results will need to be confirmed against the results indicated on the display screen. Questions are to be directed to Roche Diagnostics Technical Support at 1-800 428-2336,
- Initiated
- 2011-10-18
- Root cause
- Software design
- Terminated
- 2013-02-21
- Distribution
- Worldwide Distribution - USA including West Virginia and Texas.
- Recall number
- Z-0311-2012
- Recall status
- Terminated
- CDRH record ID
- 104528
- FDA classified
- 2011-11-30
- Recall event number
- 60146