Z-0311-2018: Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless porta

Recall record Z-0311-2018 for Philips Electronics North America Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0311-2018
Company in FDA record
Philips Electronics North America Corporation
FDA record creation date
Jan 3, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position.

Source facts

Action
If the lock mechanism of the movable clamp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is . applied. This may happen when the detector is rotated clockwise from landscape to portrait position. The Wireless Portable Detector (WPD) is fixed in the Mobile Detector Holder via a lock mechanisrn. The 4 countersunk screws, that are used to attach the lock mechanism, can become loose. If this happens and the mechanism fails, the detector may fall down when the holder is rotated 90 clockwise from Iandscape.
Initiated
2017-07-14
Root cause
Under Investigation by firm
Terminated
2020-06-05
Distribution
412 units affected.
Recall number
Z-0311-2018
Recall status
Terminated
CDRH record ID
160639
FDA record created
2018-01-03
Recall event number
78833

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗