Public FDA record

Z-0311-2025: MEGA 7.5Fr. 40cc Intra-Aortic Balloon with Accessories, Model Number 0684-00- 0295-01

Recall record Z-0311-2025 for Datascope Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0311-2025
Company in FDA record
Datascope Corp.
FDA classification date
Nov 5, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is discrepant labeling on the inner and outer packaging of the device.

Source facts

Action
An URGENT MEDICAL DEVICE REMOVAL notification letter was sent to customers beginning 9/23/24. Actions to be taken by the user: Our records indicate that you have received one or more of the affected MEGA 7.5Fr 40cc IAB devices. Please examine your inventory immediately to determine if you have any impacted product. See pictures of the outer cartons for the potentially impacted product below. Please verify all labels, IFU, and IAB against the artwork on the outer carton. If the product, IFU, and labels all match the information for the MEGA 7.5Fr 40cc outer carton (circled green in the second picture below), the device can be used as indicated in the IFU and does not need to be returned. If the outer carton states that Lot 3000344298 is a 50cc size (circled red in the first picture below), then this is incorrect and you should return the affected device by contacting Getinge QRC on QRC.UKI@getinge.com. Please forward this information to all current and potential Datascope/Getinge IAB users within your hospital/facility. If you are a distributor who has shipped any affected products to customers, please forward this letter to their attention for appropriate action. Whether or not you have any impacted product, complete and sign the attached Response Form (Page 5) to acknowledge that you have received and understand this notification. Return the completed form to Datascope/Getinge by e-mailing a scanned copy or by faxing the form to your local Datascope/Getinge Representative or office. Actions to be taken by Getinge: Datascope/Getinge is initiating this Urgent Medical Device Removal to notify users of this IAB labeling issue. This voluntary removal only affects the products identified above that are manufactured for lot number 3000344298; no other products are affected. Email: QRC.UKI@getinge.com
Initiated
2024-09-17
Root cause
Error in labeling
Distribution
International distribution to the countries of Germany, Spain , Finland, Great Britain, Hong Kong, Ireland, Italy, Lithuania, Moldova, Mexico, Norway, New Zealand, Poland, Sweden, Singapore.
Recall number
Z-0311-2025
Recall status
Open, Classified
CDRH record ID
210128
FDA classified
2024-11-05
Recall event number
95396

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗