- FDA identifier
- Z-0312-2017
- Company in FDA record
- Becton Dickinson & Co.
- FDA record creation date
- Oct 25, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
BD has determined that the PhoenixSpec AP Calibrator referenced in this recall from the following lot (5327782) are labeled with an incorrect expiration date (year). The expiration date printed on the individual calibration tubes within the kit are correct but difficult to read on a portion of the tubes.
Source facts
- Action
- All US customers were contacted via letters sent via UPS. Ex-US BD sites will contact customers in their local regions. The firm will use an acknowledgement form included with the follow up letter that is mailed or faxed back to BD will be reconciled with the customer list. BD will contact 100% of our customers that have not responded to the communication to verify acknowledgement
- Initiated
- 2016-06-01
- Root cause
- Process control
- Terminated
- 2016-11-08
- Distribution
- BD has contacted all 31 US customers and 22 Ex-US sites which received product from the associated catalog numbers. US: AL, CA, CO, KY, MD, MI, MS, NE, NJ, OH, PA, SC, SD, TX, VA, WA and WI. Ex-US: Argentina, Europe, Canada, China, Ecuador, India, Japan, New Zealand, Peru, Singapore, Thailand and Vietnam 4 US government sites were contacted
- Recall number
- Z-0312-2017
- Recall status
- Terminated
- CDRH record ID
- 150166
- FDA record created
- 2016-10-25
- Recall event number
- 75375