Open original FDA record ↗Recall record
- FDA identifier
- Z-0312-2022
- Company in FDA record
- BioFire Diagnostics, LLC
- FDA record creation date
- Nov 30, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to an increased risk of false positive Acinetobacter calcoaceticus- baumannii complex results. The cause may be as a due to non-viable contamination in the blood culture vials.
Source facts
- Action
- On or about 10/10/2021, the Recalling Firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter to U.S. customer via overnight FedEx to inform them that there is an increased risk of false positive Acinetobacter calcoaceticus- baumannii complex results when the blood culture identification panels are used with some types of blood culture vials. The increased risk are potential due to an increased level of nucleic acid from non-viable Acinetobacter calcoaceticus- baumannii complex targets in the blood culture vials. Affected Blood Culture Media vials are: 442023 - BD BACTEC Plus Aerobic medium in plastic vials; 442021 - BD BACTEC Lytic Anaerobic medium in plastic vials; 442192 - BD BACTEC Plus Aerobic medium in glass vials; 442022 - BD BACTEC Plus Anaerobic medium in plastic vials; 442020 - BD BACTEC Peds Plus mediums in plastic vials; Customers are instructed to: 1) If the BCID2 Panel is used to test BD BACTEC" blood culture vials with an expiration date of April 30, 2022 or earlier, positive results for Acinetobacter calcoaceticus- baumannii complex should be confirmed by another method prior to reporting the test results. 2) Complete the accompanying Field Action Acknowledgment of Receipt Form and return to the Recalling Firm via one of the listed return methods so that the Recalling Firm may acknowledge their receipt of this notification. Any questions or concerns, contact customer support department at Support@biofiredx.com or via telephone by dialing 1.800.735.6544 and selecting option 5 for Product Technical Support.
- Initiated
- 2021-10-11
- Root cause
- Material/Component Contamination
- Terminated
- 2023-08-10
- Distribution
- Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY. The Countries of Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bulgaria, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Dominican Rep/, Ecuador, Egypt, France, Georgia, Germany, Greece, Guam, Guatemala, Hong Kong, Hungary, India, Israel Italy, Kuwait, Latvia, Macedonia, Malaysia, Mexico, Netherlands, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Shanghai, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, Utd/Arab Emir/.
- Recall number
- Z-0312-2022
- Recall status
- Terminated
- CDRH record ID
- 190058
- FDA record created
- 2021-11-30
- Recall event number
- 88934