Open original FDA record ↗Recall record
- FDA identifier
- Z-0312-2025
- Company in FDA record
- FHC, Inc.
- FDA classification date
- Nov 8, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.
Source facts
- Action
- On October 4, 2024, the firm notified affected consignees via Urgent Medical Device Recall letters. Customers were instructed to immediately identify and quarantine all affected product. The firm will contact customers and issue an RMA to return affected product and issue a replacement by 10/11/24. Customers should not use the product in the interim. Should you have any further questions regarding this matter you may contact the firm by email at kmoeykens@fh-co.com or by telephone at 207-666-5425.
- Initiated
- 2024-10-04
- Root cause
- Process change control
- Distribution
- Domestic distribution only to the following states: AZ, CA, PA, TN, TX, WA.
- Recall number
- Z-0312-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210355
- FDA classified
- 2024-11-08
- Recall event number
- 95479