Skip to content

Public FDA record

Z-0312-2026: Brand Name: AESCULAP Product Name: REDUCING CONVERTER 10/12MM TO 5MM Model/Catalog Number: EK087P Softwar

Recall record Z-0312-2026 for Aesculap Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0312-2026
Company in FDA record
Aesculap Inc
FDA classification date
Oct 24, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was determined there is the potential of silicone fragments detaching from the yellow sealing component.

Source facts

Action
On September 24, 2025 URGENT MEDICAL DEVICE RECALL NOTIFICATION letters were sent to customers. Actions to be taken. 1. Review the recall notice and ensure all users in the organization are notified. Forward this notification to customers. Remove affected units from inventory, quarantine the product but do not destroy it. Complete and return the included response form. Return affected units to the firm. Credit of the product will be issued. An alternative disposable universal seal is available and can be ordered.
Initiated
2025-09-24
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, MD, NE, NM, NY, OR, PA, WA, WI and the country of Canada.
Recall number
Z-0312-2026
Recall status
Open, Classified
CDRH record ID
216232
FDA classified
2025-10-24
Recall event number
97649

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.