Z-0313-2025: The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative d

Recall record Z-0313-2025 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0313-2025
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Nov 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Access Cortisol assay protocol file (APF) applies an incorrect calibration curve acceptance parameter at the S1 calibrator level that can lead to a high rate of calibration failures which can vary by reagent lot number. Calibration failure would lead to delay of measurement results reporting and potential injuries.

Source facts

Action
Beckman Coulter issue and URGENT MEDICAL DEVICE RECALL notice to its consignees on 09/30/2024 via letter (USPS) and email. The notice explained the issue, the impact to the patient, and requested the use of alternative lot of Access Cortisol reagent. In the event the laboratory cannot generate a passing calibration curve use alternative methodology including alternative Beckman Coulter Access Immunoassay Systems. Beckman Coulter will release an updated Access Cortisol APF on the DxI 9000 Access Immunoassay Analyzer that includes the correct calibration curve acceptance parameter at the S1 calibrator level. A Beckman Coulter representative will contact you when the updated APF is available. The notice is to be shared with laboratory staff and forwarded to all parties to whom the the affected product was distributed.
Initiated
2024-09-30
Root cause
Process design
Distribution
Worldwide distribution.
Recall number
Z-0313-2025
Recall status
Open, Classified
CDRH record ID
210348
FDA classified
2024-11-06
Recall event number
95473

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗