Open original FDA record ↗Recall record
- FDA identifier
- Z-0314-2017
- Company in FDA record
- BBI SOLUTIONS OEM LTD
- FDA record creation date
- Oct 27, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A component of the kit is showing a sensitivity performance detected in stability testing at 15 months of an 18-month shelf life. Qualitative results may be affected. False negative results are possible which may incorrectly influence the clinical diagnostic decision.
Source facts
- Action
- BBI Solutions sent an Urgent Medical Device Recall letter dated June 30, 2016, to all affected customers. Customers were advised to immediately discontinue use of the product, consider the requirement for review of reported test results, complete and return the customer acknowledgement form to BBI Solutions and to complete and return the customer reconciliation form to the distributor. Customers who have forwarded product to other sites, should inform those customers of the product recall and provide them with a copy of the notice. Customers with questions were instructed to call Tel: +44(0)1382 564093.
- Initiated
- 2016-06-01
- Root cause
- Nonconforming Material/Component
- Terminated
- 2018-06-14
- Distribution
- US distribution to CT and IN
- Recall number
- Z-0314-2017
- Recall status
- Terminated
- CDRH record ID
- 148459
- FDA record created
- 2016-10-27
- Recall event number
- 74786