Open original FDA record ↗Recall record
- FDA identifier
- Z-0314-2027
- Company in FDA record
- Stryker Leibinger GmbH & Co. KG
- FDA classification date
- Oct 29, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient.
Source facts
- Action
- The firm began notifying customers on September 24, 2026 via Urgent Medical Device Recall letters. Customers were instructed to immediately locate and quarantine any affected products at their facility. Customers may either destroy affected products onsite with their Stryker representative and send a completed Certificate of Destruction to the firm, or return the products to Stryker. The firm will provide a replacement upon receipt of the recalled product or Certificate of Destruction.
- Initiated
- 2026-09-24
- Root cause
- Process control
- Distribution
- Worldwide - US Nationwide distribution in the states of Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Hawaii, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Michigan, Missouri, Montana, New Hampshire, New York, Ohio, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Washington and the countries of Austria, Canada, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Norway, Portugal, South Korea, Spain, Switzerland, UK.
- Recall number
- Z-0314-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222745
- FDA classified
- 2026-10-29
- Recall event number
- 99855