Open original FDA record ↗Recall record
- FDA identifier
- Z-0315-2008
- Company in FDA record
- Beckman Coulter Inc
- FDA classification date
- Dec 22, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Patient Mis-identification can occur: 1)- When a positive identifier (Sample ID or Cassette/Position) is manually edited to a positive identifier that is already in the To Do list. The workstation will accept the entry and no error message will be generated, creating 2 samples with the identical positive identifier. 2)- When manually entering a Patient ID, if a blank space is entered between th
Source facts
- Action
- Product Corrective Action (PCA) letters were mailed on April 18,2007, to all GENS, LH500, LH750 and LH780 accounts. They are informed that a potential misidentification can occur in 2 scenarios. 1. When a positive identifier (Sample ID or CassetteIPosition) is manually edited to a positive identifier that is already in the ToDo list. The workstation will accept the entry and no error message will be generated, creating 2 samples with the identical positive identifier. 2. When manually entering a Patient ID, if a blank space is entered between the characters in the Patient ID field, the system will only accept the characters before the space.----Short term fix----- The letter explains the issues and provides the customer work around. (Letters were sent by US mail). A customer response form was provided. The root cause of the problem is still under investigation, software upgrades, when approved will be made to devices.
- Initiated
- 2007-04-18
- Root cause
- Software design
- Terminated
- 2012-04-16
- Distribution
- World wide: USA and Canada
- Recall number
- Z-0315-2008
- Recall status
- Terminated
- CDRH record ID
- 64532
- FDA classified
- 2007-12-22
- Recall event number
- 44916