Intuitive Surgical has identified a software anomaly in the da Vinci Xi P5 software that can result in
unexpected master movement and potential instrument tip movement under certain circumstances
Source facts
Action
Urgent Medical Device Correction letters were sent on October 12, 2016 by traceable method to all affected customers. Letters notify customers of the issue and provided instructions to prevent it from happening until ISI software can install new software upgrade in 1st Q 2017.
Initiated
2016-10-05
Root cause
Software design
Terminated
2017-05-01
Distribution
Australia, Austria, Belgium, Colombia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Monaco, Netherlands, Norway, Portugal, Puerto Rico, Qatar, Romania, South
Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, United States
Recall number
Z-0315-2017
Recall status
Terminated
CDRH record ID
150460
FDA classified
2016-10-26
Recall event number
75473