Z-0315-2027: Carevo, Height-adjustable shower trolley, Model Number BAC1111-01

Recall record Z-0315-2027 for ARJOHUNTLEIGH POLSKA Sp. z.o.o.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0315-2027
Company in FDA record
ARJOHUNTLEIGH POLSKA Sp. z.o.o.
FDA classification date
Oct 29, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient handling. If the side support is not fully locked and an additional force is applied by the caregiver on the handle (e.g., during turning or repositioning), the panel may open unintentionally, which can lead to a resident/patient fall and serious injury, such as head injury, fractures, or other trauma resulting from a patient fall.

Source facts

Action
A FIELD SAFETY NOTICE dated 9/23/26 was sent to customers for Carevo shower trolley - side support opening by accident from outside of the device. Please review the following warnings carefully: 1. When the side panels snap into place, a characteristic "click" sound can be heard. 2. Ensure both latches on each side support are in the locked position. Failure to comply may result in unintentional side support opening while handling a patient/resident, posing their fall risk. Refer to picture 2. 3. Avoid hitting or pushing the side support handle with your thigh or knee when handling the patient. Failure to comply may result in unintentional side support opening while handling a patient, posing a fall risk. Refer to Pic.3. Unintentional opening of the side support is very unlikely if all instructions in the Instruction for Use (IFU) are followed. The locking mechanism may not return to its original position after accidental unlocking, but complete opening is unlikely if the person handling the patient is well trained and follows the IFU recommendations. Potential Risks: Failure to follow these instructions may result in patient fall, which can lead to a serious injury. Next Steps: 1. Review the enclosed Field Safety Notice and updated Instructions for Use (IFU) carefully. 2. Use the device as outlined in the IFU and perform the recommended checks. 3. Complete and return the Customer Response Form to your local Arjo representative, including details of any devices that have been sold or transferred to another facility. (Form attached to this notice) 4. Forward this alert and the updated IFU to all relevant personnel and any other organizations/persons who have received these devices. 5. Maintain awareness of this notice and the updated IFU at your site. If you have any questions or concerns regarding the safe operation of the side supports, please contact your Arjo representative. If you have any further questions or require assistance completing the Customer Respons
Initiated
2026-09-25
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-0315-2027
Recall status
Open, Classified
CDRH record ID
222844
FDA classified
2026-10-29
Recall event number
99900

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗