Open original FDA record ↗Recall record
- FDA identifier
- Z-0315-2027
- Company in FDA record
- ARJOHUNTLEIGH POLSKA Sp. z.o.o.
- FDA classification date
- Oct 29, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient handling. If the side support is not fully locked and an additional force is applied by the caregiver on the handle (e.g., during turning or repositioning), the panel may open unintentionally, which can lead to a resident/patient fall and serious injury, such as head injury, fractures, or other trauma resulting from a patient fall.
Source facts
- Action
- A FIELD SAFETY NOTICE dated 9/23/26 was sent to customers for Carevo shower trolley - side support opening by accident from outside of the device. Please review the following warnings carefully: 1. When the side panels snap into place, a characteristic "click" sound can be heard. 2. Ensure both latches on each side support are in the locked position. Failure to comply may result in unintentional side support opening while handling a patient/resident, posing their fall risk. Refer to picture 2. 3. Avoid hitting or pushing the side support handle with your thigh or knee when handling the patient. Failure to comply may result in unintentional side support opening while handling a patient, posing a fall risk. Refer to Pic.3. Unintentional opening of the side support is very unlikely if all instructions in the Instruction for Use (IFU) are followed. The locking mechanism may not return to its original position after accidental unlocking, but complete opening is unlikely if the person handling the patient is well trained and follows the IFU recommendations. Potential Risks: Failure to follow these instructions may result in patient fall, which can lead to a serious injury. Next Steps: 1. Review the enclosed Field Safety Notice and updated Instructions for Use (IFU) carefully. 2. Use the device as outlined in the IFU and perform the recommended checks. 3. Complete and return the Customer Response Form to your local Arjo representative, including details of any devices that have been sold or transferred to another facility. (Form attached to this notice) 4. Forward this alert and the updated IFU to all relevant personnel and any other organizations/persons who have received these devices. 5. Maintain awareness of this notice and the updated IFU at your site. If you have any questions or concerns regarding the safe operation of the side supports, please contact your Arjo representative. If you have any further questions or require assistance completing the Customer Respons
- Initiated
- 2026-09-25
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0315-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222844
- FDA classified
- 2026-10-29
- Recall event number
- 99900