Z-0316-2025: SONIALVISION G4 is a multi-purpose X-ray R/F system. The equipment is suitable for radiographic and fluor

Recall record Z-0316-2025 for SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0316-2025
Company in FDA record
SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION
FDA classification date
Nov 7, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may lead to the possibility of delay or cancellation of the examination.

Source facts

Action
On 9/13/24, recall notices were emailed to consignees asking them to do the following: 1) If the problem occurs it is possible to solve this problem by turning off the power of the system and then pushing the reset switch on the Digital radiography unit s control cabinet. 2) Your firm authorized service representative will be contacting to schedule the implementation of the corrective action but if you are not contacted, contact the firm's National Support. 3) Complete and return the response form via email to response to drodriguez@shimadzu-usa.com If you have questions contact the firm's National Technical Support at 844-487-2767, available 8:00AM-5:00PM, Monday - Friday.
Initiated
2024-09-06
Root cause
Software change control
Distribution
US Nationwide distribution in the states of CA, CT, MA, WV, OH, NY, IA, NE, MN, IL, FL, NC, HI, LA, MD, MT, WA, AR, TX, TN, CO.
Recall number
Z-0316-2025
Recall status
Open, Classified
CDRH record ID
210310
FDA classified
2024-11-07
Recall event number
95450

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗