Z-0316-2027: Tandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700,

Recall record Z-0316-2027 for Tandem Diabetes Care, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0316-2027
Company in FDA record
Tandem Diabetes Care, Inc.
FDA classification date
Oct 30, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class I

Reason stated in the source

Tandem Diabetes Care, Inc. ( Tandem ) has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. Investigation determined that increased friction within the drive module can reduce available motor power and, under certain conditions, cause motor speed to fall below the occlusion detection threshold. As a result, the pump may incorrectly detect an occlusion condition and suspend insulin delivery. When an occlusion alarm occurs, insulin delivery remains suspended until the user acknowledges the alarm and resumes therapy. Repeated false-positive occlusion alarms, if not addressed promptly, may result in interruptions of insulin delivery that can increase the risk of elevated blood glucose levels. In severe cases, hyperglycemia may require medical intervention or hospitalization.

Source facts

Action
On 9/23/2026, "URGENT MEDICAL DEVICE CORRECTION" recall and correction notices were emailed to customers who were asked to do the following: Correction notices: Continue to use pump while following the instructions below. 1) Always respond to an occlusion alarm and follow the recommended troubleshooting steps in the User Guide. Acknowledge the alarm, inspect the cartridge, tubing, and infusion site for signs of damage or blockage, correct any identified condition, and resume insulin delivery. 2)Ensure that your Tandem Mobi mobile app has push notifications enabled so if the alarm does occur, you will also be notified by the app. Note: If you need assistance with enabling smartphone notifications to receive alerts from your mobile app, please click the link below to access an article on firm's support page: https://www.tandemdiabetes.com/support-center/software-and-apps/tandem-tslim-mobile-app/article/enable-smartphone-notifications-to-receive-alerts 3) Contact Tandem Customer Support if: a. You are experiencing an unusually high number of occlusion alarms, b. Occlusion alarms continue to occur after following the recommended troubleshooting steps, c. Occlusion alarms unexpectedly and repeatedly disrupt your therapy, or d. You have concerns or need assistance resolving an alarm. Tandem Customer Support can review your experience and determine whether further evaluation or action is appropriate. 4) Be prepared to use a backup method of insulin delivery. 5) Complete the online acknowledgement form using this link: https://campaign.tandemdiabetes.com/FCN-Mobi-Occlusions Recall Notices informed customers to contact firm at 1-877-801-6901 or techsupport@tandemdiabetes.com to arrange exchange. Distributors were informed: If a customer contacts you regarding this field action, please direct the customer to Tandem Customer Support at 1-877-801-6901 or techsupport@tandemdiabetes.com.
Initiated
2026-09-23
Root cause
Process change control
Distribution
Worldwide - US Nationwide distribution including in the states of CA, FL, NC, OK, UT, WA, IL, PA, MN, VA, NY, IN, HI, OR, GA, MI, KS, CO, WI, SC, null, TX, NE, AZ, OH, NM, NV, MA, SD, IA, TN, ID, MD, RI, CT, NJ, ME, KY, MT, WV, VT, MS, AR, AL, DC, LA, MO, NH, PR, ND, AK, DE, WY and the country of Canada.
Recall number
Z-0316-2027
Recall status
Open, Classified
CDRH record ID
222734
FDA classified
2026-10-30
Recall event number
99851

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗