- FDA identifier
- Z-0317-2007
- Company in FDA record
- Beckman Coulter Inc
- FDA classification date
- Dec 27, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The ratio parameter assignments (numerator and denominator) of a protocol may inadvertently change after the Parameter Selection Dialog box is accessed. Issue limited only to protocols utilizing the Ratio parameter such as LeukoSure and DNA Cell Cycle.
Source facts
- Action
- A Product Corrective Action (PCA) letter was mailed on Oct 17, 2006 to FC500 MPL and MXP software accounts. They are informed that if the operator opens the Parameter Selection Dialogue box after parameters have been selected for the RATIO Parameter, the selected parameters could revert back to the default choices. If these same parameters were used in the creation of the RATIO Parameter in any other saved protocol, the protocols will also be changed once they are opened after the error occurs. An immediate user preventive action to remedy this issue each time acquisition of samples that require a protocol utilizing the ratio parameter is desired is given within the letter. Consignees are asked to complete and return the enclosed Response Form within 10 days.
- Initiated
- 2006-10-17
- Root cause
- Other
- Terminated
- 2012-02-29
- Distribution
- Nationwide and Canada
- Recall number
- Z-0317-2007
- Recall status
- Terminated
- CDRH record ID
- 49407
- FDA classified
- 2006-12-27
- Recall event number
- 36822