Skip to content

Public FDA record

Z-0317-2010: 32CM DECATHLON GOLD FUNNEL TIP STD KIT Model: DFC27SH32 15.5 Fr polyurethane twin lumen Heparin coated h

Recall record Z-0317-2010 for Spire Biomedical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0317-2010
Company in FDA record
Spire Biomedical, Inc.
FDA classification date
Nov 17, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential clinical complication is air embolism, particularly for the segment of the patient population that develops negative central venous pressure during normal respira

Source facts

Action
Spire Medical issued an "Urgent device Recall" letter dated 10/22/09 by FEDEX to end users. Users are requested to return product. For questions, contact Spire Biomedical Customer Service at 1-877-876-8635.
Initiated
2009-10-23
Root cause
Process control
Terminated
2010-06-09
Distribution
Worldwide distribution - Nationwide, Government. Foreign: Canada, France.
Recall number
Z-0317-2010
Recall status
Terminated
CDRH record ID
86001
FDA classified
2009-11-17
Recall event number
53605

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗