Z-0317-2027: Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time (APTT) Test Cuv

Recall record Z-0317-2027 for Accriva Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0317-2027
Company in FDA record
Accriva Diagnostics, Inc.
FDA classification date
Oct 30, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Microcoagulation test used to evaluate low dose heparin anticoagulation has a mispositioned or absent clot activating reagent, which may result in out-of-range high error code, or erroneously high results, which may lead retest, or overestimation of patient's anticoagulation, leading to inadequate anticoagulant therapy, which could result in thrombosis.

Source facts

Action
On 9/14/2026, recall notices were mailed to customers who were informed of the following: 1) Please forward this communication to those within your organization who need to be aware of this matter and verify they understand the procedure and can complete the enclosed forms. 2) Forward the enclosed customer notification to all other organizations where affected devices may have been transferred to. 3) Please check your inventory and do not use affected lots. 4) Scrap/destroy all impacted cuvettes and/or boxes. 5) Complete and return the Acknowledgement and Receipt Form via email to ilsd_ra.na@werfen.com Questions about receiving credit, should be directed to Customer Service at 1-800-955-9525 or e-mail customerservice.na@werfen.com. Technical questions should be directed to Technical Services at 1-800-678-0710 Option 3 or email techsupport.na@werfen.com.
Initiated
2026-09-14
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution in the states of MA, MO, AK, WV, NJ, TX, VA, IL, NY and the countries of China, Colombia.
Recall number
Z-0317-2027
Recall status
Open, Classified
CDRH record ID
222749
FDA classified
2026-10-30
Recall event number
99863

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗