Reason stated in the source
The CA 125 assay product inserts list the incorrect storage instructions for the CA 125 Calibrators and Controls, and the product information letter (PI-P) shipped with the products to correct the storage conditions, 69-6349/R2, refers to an outdated control package insert, 39-1642/R3, instead of the current control package insert 34-1047/R4
Source facts
- Action
- Field Correction recall by letter dated 11/25/03 to all customers receiving the affected reagents. The accounts were informed of the incorrect control package insert referenced in the product information letter shipped with each CA 125 product. The letters contained the correct control package insert number, 34-1047/R4, and stated that all CA 125 assay product inserts will be updated shortly to include the revised storage instructions for the CA 125 Calibrators and Controls.
- Initiated
- 2003-11-25
- Root cause
- Other
- Terminated
- 2004-07-26
- Distribution
- Nationwide and internationally through Abbott subsidiaries in Argentina, Aruba, Australia, Bermuda, Brazil, Canada, Cayman Islands, Chile, Colombia, Ecuador, Germany, Peru, Hong Kong, India, Jamaica, Japan, Korea, New Zealand, Pakistan, Puerto Rico, Singapore, Taiwan, Thailand, United Kingdom and Uruguay.
- Recall number
- Z-0318-04
- Recall status
- Terminated
- CDRH record ID
- 30604
- FDA classified
- 2004-01-10
- Recall event number
- 27834