Open original FDA record ↗Recall record
- FDA identifier
- Z-0318-2022
- Company in FDA record
- Magnolia Medical Technologies, Inc.
- FDA record creation date
- Dec 1, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to an uptick of complaints associated with premature needle retraction of blood collection system.
Source facts
- Action
- On 11/02/2021, the Recalling Firm sent an "URGENT: MEDICAL DEVICE RECALL" Notification Letter to customers via email informing them that the firm has received a number of product complaints associated with its blood collecting systems where the retractable feature of the butterfly needles may prematurely activate. The product complaints have been associated with retractable feature being activated within the product packaging. Customers are instructed to: -Determine the quantity of the affected products remaining at their facility and communicated this to the Recalling Firm's Sales or Clinical team members. -The Recalling Firm will replace any of the affected products at no cost to the customer. -Notify all their staff of the recall and if the affected products have been further distributed, to contact/notify those customers/organization of this recall.
- Initiated
- 2021-11-02
- Root cause
- Process control
- Terminated
- 2023-05-08
- Distribution
- U.S. Nationwide distribution in the states of MD, WV, DE, VA, DC, IN, MI, MA, SC, NY, IL, TX, AL, OK, GA, NE, FL, MO, CA, CO, NM, AR, WY, KS, PA, WA, MT, and AZ.
- Recall number
- Z-0318-2022
- Recall status
- Terminated
- CDRH record ID
- 190017
- FDA record created
- 2021-12-01
- Recall event number
- 88919