Skip to content

Public FDA record

Z-0318-2022: The Steripath Gen2 Blood Collection System Model Number/Description 2700-21-EN / Steripath Gen2 Blood Co

Recall record Z-0318-2022 for Magnolia Medical Technologies, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0318-2022
Company in FDA record
Magnolia Medical Technologies, Inc.
FDA record creation date
Dec 1, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to an uptick of complaints associated with premature needle retraction of blood collection system.

Source facts

Action
On 11/02/2021, the Recalling Firm sent an "URGENT: MEDICAL DEVICE RECALL" Notification Letter to customers via email informing them that the firm has received a number of product complaints associated with its blood collecting systems where the retractable feature of the butterfly needles may prematurely activate. The product complaints have been associated with retractable feature being activated within the product packaging. Customers are instructed to: -Determine the quantity of the affected products remaining at their facility and communicated this to the Recalling Firm's Sales or Clinical team members. -The Recalling Firm will replace any of the affected products at no cost to the customer. -Notify all their staff of the recall and if the affected products have been further distributed, to contact/notify those customers/organization of this recall.
Initiated
2021-11-02
Root cause
Process control
Terminated
2023-05-08
Distribution
U.S. Nationwide distribution in the states of MD, WV, DE, VA, DC, IN, MI, MA, SC, NY, IL, TX, AL, OK, GA, NE, FL, MO, CA, CO, NM, AR, WY, KS, PA, WA, MT, and AZ.
Recall number
Z-0318-2022
Recall status
Terminated
CDRH record ID
190017
FDA record created
2021-12-01
Recall event number
88919

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗