Skip to content

Public FDA record

Z-0318-2026: Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US

Recall record Z-0318-2026 for Dexcom, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0318-2026
Company in FDA record
Dexcom, Inc.
FDA classification date
Oct 30, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could result in the missed detection of a hyperglycemic or hypoglycemic event, protentional resulting in severe hyperglycemia, diabetic ketoacidosis (DKA), or hyperosmolar hyperglycemic state (HHS).

Source facts

Action
On 09/18/2025, the firm sent via in-app and email an "URGENT: MEDICAL DEVICE CORRECTION NOTIFICATION" communication informing customer that a bug has been identified within the G6 Android app version 1.15.0 that can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, or alerts. Customers are instructed to: Upgrading from app version 1.15.0 is mandatory and they will not be able to use app version 1.15.0 after October 6th, 2025. Update your app now to continue using the app: 1. Open the App Store and search for the Dexcom G6 app. 2. Install the latest Dexcom G6 app version. 3. Open the Dexcom G6 app. For questions or further assistance - contact Technical Support at 1-844-478-1600. Technical Support is available 24 hours a day/7 days a week
Initiated
2025-08-28
Root cause
Software Design Change
Distribution
This product was distributed throughout the US. G6 Android US v1.15.0 SW11678 G6 and G6 Pro Android app release has been halted <1% rollout after this issue was identified. Please see the attached consignee data for additional details.
Recall number
Z-0318-2026
Recall status
Open, Classified
CDRH record ID
216085
FDA classified
2025-10-30
Recall event number
97607

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗