Mindray DS USA, Inc., dba Datascope Patient Monitoring
FDA classification date
Dec 5, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Two issues have been identified with the AS3000 unit.
1. The threadlocker of the caster (wheels) may not have been utilized, which may allow the caster to loosen and possibly separate from the unit.
2. The use of select brands of pre-pack absorber in the absorber canister of the AS3000 has been associated, in some cases, with gas leakage around the pre-pack, rather than through the absorber mate
Source facts
Action
Notification (Urgent Product Field Correction) letters were sent to the one consignee on October 7, 2008 by Federal Express overnight delivery.
Initiated
2008-10-07
Root cause
Component design/selection
Terminated
2008-12-08
Distribution
Only 2 units are in distribution, both in the same facility in Redondo Beach, CA.
Recall number
Z-0319-2009
Recall status
Terminated
CDRH record ID
74064
FDA classified
2008-12-05
Recall event number
49713
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.