Skip to content

Public FDA record

Z-0319-2021: Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Sup

Recall record Z-0319-2021 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0319-2021
Company in FDA record
Boston Scientific Corporation
FDA record creation date
Oct 31, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is potential when a user changes an EGM trace channel with a Manual Pace Threshold test in progress that the newly modified trace channel may change to a flat line.

Source facts

Action
Boston Scientific sales representatives will hand-deliver a removal letter to all customer consignees with one of the eight affected Model 3300 LATITUDE programmer. BSC sales representatives will maintain control of these devices. In order to continue to provide the benefits that this software support application provides for those patients implanted with an ALTRUA/INSIGNIA I pacemaker, BSC sales representatives will continue to use Model 3300 LATITUDE programmers installed with the Model 3892 v1.04 ALTRUA / INSIGNIA I / NEXUS I Software Support Application in the field. The risks associated with this unintended behavior can be avoided when specific remedial actions are adhered to. These remedial actions will be adhered to until a software correction is approved and installed on these programmers. BSC will confirm via email from sales representatives that 100% of the removal letters were delivered. BSC will post the removal letter on the BSC website within the Product Performance Resource Center located @ www.bostonscientific.com/en-US/pprc/productadvisories.html.
Initiated
2020-09-22
Root cause
Software design
Terminated
2024-09-04
Distribution
Distributed in US - AR, FL, IL, and TX
Recall number
Z-0319-2021
Recall status
Terminated
CDRH record ID
183949
FDA record created
2020-10-31
Recall event number
86534

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗