Skip to content

Public FDA record

Z-0319-2025: Coated VICRYL Polyglactin 910 Suture VIO 36IN(90CM) USP0(M3.5) S/A CT, Product Number J358H

Recall record Z-0319-2025 for Ethicon, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0319-2025
Company in FDA record
Ethicon, Inc.
FDA classification date
Nov 7, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Complaints were received regarding needle pull-off and suture degradation. Analysis on returned product from this lot confirmed that some sutures were visually degraded in the package.

Source facts

Action
An URGENT: FIELD SAFETY NOTICE was sent to customers beginning on 9/25/24. ACTION REQUIRED 1. Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). If you have product subject to this recall, please maintain a copy of this notice with the quarantined product and keep a copy for your records. 2. Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 3. Complete the Business Reply Form (BRF) (Attachment 3) confirming receipt of this notice and fax or email to [Enter Affiliate Information] within three (3) business days. Please return the BRF even if you do not have product subject to this recall. 4. Customers are required to return unused VICRYL" (polyglactin 910) Suture subject to this recall that are in inventory immediately. To receive credit or replacement product as appropriate, customers must return product subject to this recall no later than December 31, 2024 to [Enter Affiliate Information]. Any non-affected product and any product returned after the date specified will not be eligible for credit or replacement product. 5. Keep this notice visibly posted for awareness until all product subject to this recall has been returned to [Enter Affiliate Information]. If you require any assistance with
Initiated
2024-09-25
Root cause
Under Investigation by firm
Distribution
International distribution to the country of China.
Recall number
Z-0319-2025
Recall status
Open, Classified
CDRH record ID
210476
FDA classified
2024-11-07
Recall event number
95516

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗