Open original FDA record ↗Recall record
- FDA identifier
- Z-0319-2027
- Company in FDA record
- NATUS MEDICAL INCORPORATED
- FDA classification date
- Oct 30, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
The device may be unable to correctly format its internal memory. When the internal memory cannot be correctly formatted, previously recorded studies may remain on the device and may be updated with incorrect patient information from a subsequent study. The issue may also prevent studies from initiating recording successfully or may cause recordings to end prematurely, resulting in incomplete or unavailable diagnostic data and the potential need to repeat a patient study.
Source facts
- Action
- An Urgent Medical Device Field Correction notification letter was sent to customers beginning 9/29/26. ACTION: " Review the serial numbers in this notice and identify any affected devices that are currently in your possession. " Immediately cease use of the affected product and quarantine until you receive further instructions from Natus for correcting the affected device(s). " Review all studies recorded on affected Embletta MPR devices to verify that patient information is correctly associated with the corresponding study and that the study data are complete, accurate, and attributable to the intended patient. " Please complete the enclosed form and return to FCA@Natus.com. If you have affected product, you will receive instructions on how to correct your affected Embletta MPR PG device(s) from FCA@Natus.com. All corrections will be provided at no cost to you. " If you do not have affected product, please respond to confirm no further actions are required. " Adverse reactions or quality problems experienced with the use of the part should be reported to Natus and the FDA's MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. " Complete and submit the report Online at www.fda.gov/medwatch/report.htm. " Regular Mail or Fax: Download form or call 1- 800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178. RESOLUTION Investigation results have determined that the issue is caused by a recent change to the internal memory within the affected Embletta MPR devices. The corrective action is currently being finalized. Customers with affected devices should contact Natus using the information provided above. Natus will provide the necessary instructions to complete the corrective action. If you have any questions regarding this notice, please contact FCA@Natus.com.
- Initiated
- 2026-09-29
- Root cause
- Component change control
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, China, Italy, Portugal.
- Recall number
- Z-0319-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222948
- FDA classified
- 2026-10-30
- Recall event number
- 99941