Open original FDA record ↗Recall record
- FDA identifier
- Z-0321-2024
- Company in FDA record
- Medtronic Inc
- FDA classification date
- Nov 17, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Mo.Ma Ultra device manifold is labelled with 2 stickers, an inferior and superior sticker, one on each side of the manifold. These stickers label the common carotid artery (CCA) proximal balloon and external carotid artery (ECA) distal balloon inflation and deflation ports. Both sides of the complaint units manifold were labelled with the superior label leading to an incorrect identification of the proximal and distal inflation /deflation ports when viewing the devices from the inferior side.
Source facts
- Action
- Medtronic issued an URGENT: MEDICAL DEVICE RECALL NOTIFICATION on 10/04/2023 via UPS. The notice explained the issue, potential risk, and requested the return of the device. For questions, consignees were directed to contact their Medtronic Field Sales Representative.
- Initiated
- 2023-10-12
- Root cause
- Process control
- Distribution
- Worldwide distribution - US Nationwide and the countries of Algeria, Armenia, France, Germany, Iraq, Italy, Japan, Netherlands, Poland, Romania, Spain, Sweden, Switzerland, United Kingdom.
- Recall number
- Z-0321-2024
- Recall status
- Open, Classified
- CDRH record ID
- 203779
- FDA classified
- 2023-11-17
- Recall event number
- 93258