Open original FDA record ↗Recall record
- FDA identifier
- Z-0322-2018
- Company in FDA record
- CareFusion 303, Inc.
- FDA record creation date
- Jan 8, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The syringe plunger grippers may fail to automatically close around the syringe plunger press when the gripper control knob has been closed.
Source facts
- Action
- The recalling firm issued "Medical Device Safety Notification, dated November 17, 2017, via UPS. The letter describes the potential issue, affected products, potential risks, and required action for the users. The actions noted that clinicians can continue to use the device and follow the Alaris System User Manual for proper loading of the syringe. Per the manual, clinicians should ensure the plunger grippers lock and hold the syringe plunger in place. The Biomedical Engineering can detect if the device has this issue and if the issue described occurs, the user is to contact BD Support Center or email them to schedule service of the device at the BD Service Depot. Enclosed with the notification letter was a list of affected serial numbers, an excerpt from the user manual to "Prepare Syringe and Administration Set (Syringe Module)", and a customer response card to be returned via mail, email or FAX to confirm the notification has been received. For further questions, please call (858) 617-5977.
- Initiated
- 2017-11-17
- Root cause
- Nonconforming Material/Component
- Terminated
- 2020-05-04
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution. There was also government and military distribution. Foreign distribution was made to Canada, United Arab Emirates, Bahrain, Canada, Switzerland, United Kingdom, Kuwait, Mexico, Malaysia, The Netherlands, New Zealand, Qatar, Saudi Arabia, Taiwan, and South Africa.
- Recall number
- Z-0322-2018
- Recall status
- Terminated
- CDRH record ID
- 159839
- FDA record created
- 2018-01-08
- Recall event number
- 78588