Failure to operate: The unit display may go blank and reset when the start button is pressed, due to an internal electrical short, resulting in a centrifugal pump that fails to start as expected.
Source facts
Action
Consignees were notified via an Urgent Medical Device Correction letter dated 9/10/07, given instructions to follow if the problem were to manifest itself, and informed the firm would contact them to schedule an upgrade.
Initiated
2007-09-10
Root cause
Component change control
Terminated
2009-07-07
Distribution
Nationwide.
Recall number
Z-0323-2008
Recall status
Terminated
CDRH record ID
53816
FDA classified
2007-12-01
Recall event number
38467