Skip to content

Public FDA record

Z-0323-2017: Visualase Body Accessory Kit PN 020-3000. The individual components within the Visualase Body Accessory K

Recall record Z-0323-2017 for Medtronic Navigation, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0323-2017
Company in FDA record
Medtronic Navigation, Inc.
FDA classification date
Oct 28, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field.

Source facts

Action
Medtronic sent an Urgent Medical Device Recall letter dated September 2, 2016, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to examine their stock for the affected product and if any are found contact Medtronic to arrange for a no charge replacement. Customers were asked to complete the Customer Confirmation form attached to the letter and return it to Medtronic at RS.NavFCA@medtronic.com or by Fax to 651-367-7075. Customers with questions were instructed to contact Medtronic Technical Services at 1-800-595-9709. For questions regarding this recall call 720-890-3200.
Initiated
2016-09-01
Root cause
Mixed-up of materials/components
Terminated
2017-03-27
Distribution
Worldwide Distribution - US including CA, GA, FL, and TX. Internationally to Australia.
Recall number
Z-0323-2017
Recall status
Terminated
CDRH record ID
149419
FDA classified
2016-10-28
Recall event number
75124

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗