Open original FDA record ↗Recall record
- FDA identifier
- Z-0323-2017
- Company in FDA record
- Medtronic Navigation, Inc.
- FDA classification date
- Oct 28, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field.
Source facts
- Action
- Medtronic sent an Urgent Medical Device Recall letter dated September 2, 2016, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to examine their stock for the affected product and if any are found contact Medtronic to arrange for a no charge replacement. Customers were asked to complete the Customer Confirmation form attached to the letter and return it to Medtronic at RS.NavFCA@medtronic.com or by Fax to 651-367-7075. Customers with questions were instructed to contact Medtronic Technical Services at 1-800-595-9709. For questions regarding this recall call 720-890-3200.
- Initiated
- 2016-09-01
- Root cause
- Mixed-up of materials/components
- Terminated
- 2017-03-27
- Distribution
- Worldwide Distribution - US including CA, GA, FL, and TX. Internationally to Australia.
- Recall number
- Z-0323-2017
- Recall status
- Terminated
- CDRH record ID
- 149419
- FDA classified
- 2016-10-28
- Recall event number
- 75124