Open original FDA record ↗Recall record
- FDA identifier
- Z-0323-2022
- Company in FDA record
- Boston Scientific Corporation
- FDA record creation date
- Dec 1, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Balloon pinhole results in losss pressure or fails to gain or maintain pressure, causing prolonged procedure
Source facts
- Action
- Boston Scientific issued Urgent Medical Device Removal Letter to US consignees via Federal Express Priority Overnight mail on October 21, 2021. Letter states reason for recall, health risk and action to take: Cease further distribution or use of any remaining product affected by this removal immediately. Verify by product lot/batch number in the product table whether any product within your inventory is affected. If so, indicate on your Reply Verification Tracking Form the quantity of units from each lot/batch that you will be returning. If you identify any product from the affected lots/batches within your inventory, please segregate the product immediately and return it to BSC in accordance with the enclosed instructions. If you are a distributor, please note that the depth of this removal is to the end user level and this notification should be forwarded to your customers. If you are a facility that has sent products to another hospital or facility within your network, please ensure that this notification is forwarded to them. Please read carefully through the removal instructions included with this notification. Your local Sales Representative can answer any questions that you may have regarding this notification.
- Initiated
- 2021-10-21
- Root cause
- Device Design
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0323-2022
- Recall status
- Open, Classified
- CDRH record ID
- 190212
- FDA record created
- 2021-12-01
- Recall event number
- 88991