Open original FDA record ↗Recall record
- FDA identifier
- Z-0324-2008
- Company in FDA record
- Mini-Mitter Company, Inc.
- FDA classification date
- Jan 17, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Communication Failure: the devices have a memory chip that has a limited number of read/write cycles which can trigger set up information and/or subject information anomalies or intermittent communication annomalies. If left unaddressed, the issues can lead to permanent communication failure.
Source facts
- Action
- A Voluntary Medical Device Recall letter dated June 26, 2007. The recall letter will instruct the customers to 1) fill out a business reply form to confirm contact information and devices serial numbers in their possession. Then FAX this info back immediately to Stericycle or Mini-Mitter. 2) Refer to the work scheduled rework date reserved for them and noted in their recall notification letter and return the device for service at that time; and 3) Package the device appropriately, insure it and return it via UPS using the provided UPS shipping label. Product returned to Mini-Mitter will be quarantined, decontaminated, and repaired. Min-Mitter will rework all the devices returned. Each will have a new memory chip installed, be re-calibrated, retested, and have a new battery installed if required. Min-Mitter will provide a new one year warranty.
- Initiated
- 2007-06-27
- Root cause
- Software design
- Terminated
- 2010-10-20
- Distribution
- Worldwide
- Recall number
- Z-0324-2008
- Recall status
- Terminated
- CDRH record ID
- 53493
- FDA classified
- 2008-01-17
- Recall event number
- 38368