It has been discovered that during Empowr PS Total Knee replacement, while chiseling for bone removal in the box area, femoral chipping has occurred in surgery. The current design of the captured box cut guides can create an instance where the surgeon is not fully completing the box cut on the side.
Source facts
Action
The recalling firm sent a recall notification letter to affected consignees on 9/19/16. The field safety notice provided a newly revised Surgical Technique to educate users on proper use of the device.
Initiated
2016-09-19
Root cause
Under Investigation by firm
Terminated
2017-10-20
Distribution
US
Recall number
Z-0325-2017
Recall status
Terminated
CDRH record ID
149677
FDA record created
2016-10-28
Recall event number
75214