There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.
Source facts
Action
Baxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 10/11/2024 via USPS First Class Mail. The notice explained the issue, potential hazard to the patient, and provided instructions for the to the system IT administrator to correct the issue which is already available in the current version of software.
Initiated
2024-10-11
Root cause
Software design
Distribution
Worldwide distribution.
Recall number
Z-0325-2025
Recall status
Open, Classified
CDRH record ID
210594
FDA classified
2024-11-07
Recall event number
95559