Skip to content

Public FDA record

Z-0325-2027: CryoPatch SynerGraft (SG) Pulmonary Human Cardiac Patch, Model: SGP010

Recall record Z-0325-2027 for Artivion, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0325-2027
Company in FDA record
Artivion, Inc
FDA classification date
Oct 30, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

A cryopreserved pulmonary human cardiac patch, taken from a 64-year-old human donor, was outside the cardiac tissue donor age specification of newborn to 60 years. The patch was implanted, and due to limitations of long-term in vivo data, the possibility of premature graft dysfunction cannot be excluded.

Source facts

Action
On 9/18/2026, the notification was delivered via mail to customer who was advised of the following: Available biomechanical and histological evidence does not indicate an increased risk that would warrant additional patient testing, enhanced clinical monitoring, or other clinical intervention beyond the patient's routine standard of care. Accordingly, prophylactic removal or surgical replacement of the already implanted tissue solely because of the donor-age nonconformance is not clinically warranted. Complete and return the response form via email to fieldassurance@artivion.com Firm requests that any observed deficiency, performance concern, or other issue potentially associated with the tissue be reported to them vial email: fieldassurance@artivion.com or phone: +1-770-419-3355
Initiated
2026-09-18
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution in the state of IL.
Recall number
Z-0325-2027
Recall status
Open, Classified
CDRH record ID
223129
FDA classified
2026-10-30
Recall event number
100012

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗