There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential clinical complication is air embolism, particularly for the segment of the patient population that develops negative central venous pressure during normal respira
Source facts
Action
Spire Medical issued an "Urgent device Recall" letter dated 10/22/09 by FEDEX to end users. Users are requested to return product. For questions, contact Spire Biomedical Customer Service at 1-877-876-8635.
Initiated
2009-10-23
Root cause
Process control
Terminated
2010-06-09
Distribution
Worldwide distribution - Nationwide, Government. Foreign: Canada, France.
Recall number
Z-0326-2010
Recall status
Terminated
CDRH record ID
86010
FDA classified
2009-11-17
Recall event number
53605