Skip to content

Public FDA record

Z-0326-2017: Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling installed rail systems: Multira

Recall record Z-0326-2017 for Hill-Rom, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0326-2017
Company in FDA record
Hill-Rom, Inc.
FDA classification date
Oct 31, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ceiling Bracket 71 was shipped with a larger center hole. A potentially hazardous situation exists that if the Ceiling Bracket 71 with a larger center hole is used for a concrete ceiling Ultra System installation with the M8 screw and the safe working load is exceeded, the system could fall.

Source facts

Action
On 8/9/2016, URGENT: MEDICAL DEVICE CORRECTION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If you have any questions regarding this Field Safety Correction, please contact Hill-Rom Technical Support at 1-800-445-3720, your distributor or your Hill-Rom representative.
Initiated
2016-08-09
Root cause
Mixed-up of materials/components
Terminated
2017-03-24
Distribution
WA NC OH PA MI Australia Sweden
Recall number
Z-0326-2017
Recall status
Terminated
CDRH record ID
149635
FDA classified
2016-10-31
Recall event number
75197

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗