Open original FDA record ↗Recall record
- FDA identifier
- Z-0326-2017
- Company in FDA record
- Hill-Rom, Inc.
- FDA classification date
- Oct 31, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ceiling Bracket 71 was shipped with a larger center hole. A potentially hazardous situation exists that if the Ceiling Bracket 71 with a larger center hole is used for a concrete ceiling Ultra System installation with the M8 screw and the safe working load is exceeded, the system could fall.
Source facts
- Action
- On 8/9/2016, URGENT: MEDICAL DEVICE CORRECTION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If you have any questions regarding this Field Safety Correction, please contact Hill-Rom Technical Support at 1-800-445-3720, your distributor or your Hill-Rom representative.
- Initiated
- 2016-08-09
- Root cause
- Mixed-up of materials/components
- Terminated
- 2017-03-24
- Distribution
- WA NC OH PA MI Australia Sweden
- Recall number
- Z-0326-2017
- Recall status
- Terminated
- CDRH record ID
- 149635
- FDA classified
- 2016-10-31
- Recall event number
- 75197