Z-0326-2023: Busse SPINE INJECTION TRAY-Intended as a nerve block REF No. 8563R1

Recall record Z-0326-2023 for Busse Hospital Disposables, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0326-2023
Company in FDA record
Busse Hospital Disposables, Inc.
FDA classification date
Nov 29, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Kits containing 3M Steri-Drape Surgical Drapes recalled by 3M due to liner on the adhesive component is difficult to remove without damaging the product and may render the product unusable

Source facts

Action
Busse issued Product Recall Notification Letter dated 10/20/22 to Distributor; End User via Email on 10/21/2022 and Certified Mail on 10/24/2022 Letter states reason for recall, health risk and action to take: 1. Immediately check your inventory for the lots listed above. 2. Do not use the drape included in the product with the lot number(s) listed above. 3. Complete the enclosed Response Form within the next 10 days and return via email to csabillon@busseinc.com even if you have no affected product on hand. a. If you do have product in stock, destroy the drape following your facility policy.
Initiated
2022-10-21
Root cause
Device Design
Distribution
TX
Recall number
Z-0326-2023
Recall status
Open, Classified
CDRH record ID
196746
FDA classified
2022-11-29
Recall event number
91079

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗