Skip to content

Public FDA record

Z-0326-2027: Senographe Pristina

Recall record Z-0326-2027 for GE Medical Systems, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0326-2027
Company in FDA record
GE Medical Systems, LLC
FDA classification date
Nov 4, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).

Source facts

Action
GE HealthCare will send a customer letter dated August 28, 2026, that provides safety instructions for immediate mitigation of the incorrect configuration of the collimator anode track / filter in manual mode during 3D+2D acquisitions. GE HealthCare will correct all affected Senographe Pristina VIA systems by upgrading the software at no cost to the customer. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2026-08-28
Root cause
Radiation Control for Health and Safety Act
Distribution
USA and OUS.
Recall number
Z-0326-2027
Recall status
Open, Classified
CDRH record ID
223255
FDA classified
2026-11-04
Recall event number
100070

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗