Skip to content

Public FDA record

Z-0327-2016: NxStage System One S Cycler (High Permeability Hemodialysis System) Model no. NX1000-5-A

Recall record Z-0327-2016 for NxStage Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0327-2016
Company in FDA record
NxStage Medical, Inc.
FDA record creation date
Nov 23, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ultrafiltration (UF) Volume software error inaccurate fluid removal

Source facts

Action
NxStage issued issued a Field Safety Notice (FSN) on 10/29/15 to consignees who have NX1000-5 or a NX1000-5-A cycler. The FSN included Instructions for steps to be taken so that the cycler and treatments are not affected by the software error. NxStage will be sending a representative to the facility to attach detailed instructions on each cycler as a user reference. Return the attached reply form to acknowledge receipt of this letter. Questions or comments, please contact NxStage Customer Service at 1-866-NXSTAGE (1-866-697-8243).
Initiated
2015-10-29
Root cause
Software Design Change
Terminated
2019-04-29
Distribution
Worldwide Distribution. US Nationwide, Netherlands, and Sweden.
Recall number
Z-0327-2016
Recall status
Terminated
CDRH record ID
141491
FDA record created
2015-11-23
Recall event number
72543

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗