Z-0327-2025: Baxter AK 98 Dialysis Machine, Hemodialysis Delivery System, Product Code 955607. As of February 2025, t

Recall record Z-0327-2025 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0327-2025
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Nov 13, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Baxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.

Source facts

Action
Baxter sent an IMPORTANT MEDICAL DEVICE CORRECTION notice to its consignees on 10/16/2024 via USPS. Until Baxter has transitioned to newly manufactured platinum-cured silicone tubing components in AK 98 machines, the firm recommends these actions: - If you have new AK 98 machines with a runtime less than 500 hours, consider using other available machines including AK 98 machines with a runtime greater than 500 hours, as PCB and PCBA levels decrease over time of use. - If no alternative is available, continue to use the AK 98 to ensure your patients have access to dialysis treatment. - Do not stop dialysis treatment to patients who need it. - Call Baxter Renal Customer Care at 800-525-2623, option 2, option 2, between 7:00 am and 5:00 pm CT, Monday thru Friday to provide the serial numbers of your AK 98 machines with a runtime less than 500 hours to determine a correction plan for your site. - Until the solution is implemented to reduce the possible presence of PCB and PCBA below the FDA recommended limits, Baxter is holding shipment of new AK 98 machines, allowing for the treatment needs of the dialysis population in our customers care. If you have a medical need for a new AK 98 machine with the peroxide-cured silicone tubing, contact your Baxter Renal Sales Representative for instructions on placing the order. - Please forward a copy of this communication to other facilities or departments that have this product. ***Updated 12/3/25*** On 12/3/2025, Vantive sent a follow-up Important Medical Device Correction letter. This letter contained an update to health hazard based on Vantive's toxicological risk assessment. The issue is unlikely to pose a risk for patients with a body weight greater than 25kg, irrespective of the run-time of the machine. The firm has determined that no corrections are needed for devices currently with customers, and patients may continue therapy with the current configuration of AK 98.
Initiated
2024-10-16
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution and worldwide
Recall number
Z-0327-2025
Recall status
Open, Classified
CDRH record ID
210544
FDA classified
2024-11-13
Recall event number
95527

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗