Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.
Source facts
Action
Cannon Medical began notifying consignees via email on about 10/16/2025. Consignees were instructed that a representative of Canon Medical Systems USA will contact them to schedule a time to replace the relevant parts. If the message "Gradient power supply error is displayed before the replacement is performed, immediately stop using the system and contact your service representative. Consignees were requested to complete and return the provided response form
Initiated
2025-10-16
Root cause
Device Design
Distribution
US distribution to Ohio and Nevada
Recall number
Z-0327-2026
Recall status
Open, Classified
CDRH record ID
216466
FDA classified
2025-10-24
Recall event number
97735