The firm has received 42 complaints regarding the head end lock rod, head end slide tube, brushing, and upper cross tube failures.
Source facts
Action
Stryker Medical sent an Urgent Medical Device Alert letter dated May 15, 2010 to all customers. The letter identified the product, the problem, and the action to be taken by the customer. The letter stated that there is a possibility for fatigue related failures of certain components of the cot and that a Stryker representative will visit the firm to correct the device.
A representative from Stryker field service would contact the customer to perform the device corrections, free of charge.
Immediate Action required by the customer:
Locate the ambulance cots listed in this notice.
Return the enclosed post card to confirm receipt of this notification.
If any of the cots listed in the letter were loaned or sold the customer was to forward a copy of the notice to the new users and advise Stryker of their new location.
If the customer disposed of any of the affected stretchers and they were no longer in use, customers were to advise Stryker of their obsolescence by providing the serial numbers
For any questions or comments regarding this issue, please call (800) 869-0770, ext. 6902.
Initiated
2010-05-15
Root cause
Device Design
Terminated
2012-02-07
Distribution
Worldlwide Distribution - US including AZ, CA, CO, FL, LA, MS, NC, NJ, OH, PA, and TN and the countries of Australia, China, France, Germany, Hong Kong, Italy, Japan, New South Wales, Poland, Portugal, Spain, Sweden, Switzerland, and the United Kingdom.
Recall number
Z-0328-2011
Recall status
Terminated
CDRH record ID
94146
FDA classified
2010-11-08
Recall event number
56628