Open original FDA record ↗Recall record
- FDA identifier
- Z-0328-2012
- Company in FDA record
- Varian Medical Systems, Inc.
- FDA classification date
- Dec 1, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Plastic needle tip could possibly crack or separate.
Source facts
- Action
- Varian Medical Systems sent a "CUSTOMER TECHNICAL BULLETIN" letter dated September 30, 2008 to all affected customers. The letter described the product, problem, and actions to be taken by the customers. The letter instructed users to inspect needle tips before and after every clinical application to prevent needle tip separation or cracking.
- Initiated
- 2008-09-30
- Root cause
- Nonconforming Material/Component
- Terminated
- 2011-12-07
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of Arizonia, California, Florida, Illinois, Indiana, Michigan, North Carolina, and Vermont, and the countries of Argentina, Austria, Belgium, Canada, Estonia, France, Germany, India, Norway, Pakistan, Russia, Slovenia, So. Africia, Spain, Switzerland, United Kingdom, and Ukraine.
- Recall number
- Z-0328-2012
- Recall status
- Terminated
- CDRH record ID
- 103821
- FDA classified
- 2011-12-01
- Recall event number
- 59779