Open original FDA record ↗Recall record
- FDA identifier
- Z-0328-2017
- Company in FDA record
- SentreHeart Inc
- FDA classification date
- Nov 9, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
A complaint was received on September 27th, 2016 at a medical center in Germany for delamination and flaking of the PTFE lubricious coating on three FindrWIRZ Guidewire System .025 guidewires removed from their packaging hoops.
Source facts
- Action
- SentreHeart sent an Urgent Medical Device Voluntary Recall letter dated September 29, 2016, to all affected customers by traceable delivery method. Letters explained the issue and requested that any devices from the two affected lots be returned to SentreHEART immediately. Customers were asked to complete and return the enclosed response form as soon as possible via fax at 650-354-1204 or email to customerservice@sentreheart.com. Customers with questions were instructed to call 1-855-256-7330. For questions regarding this recall call 650-241-6008.
- Initiated
- 2016-09-29
- Root cause
- Under Investigation by firm
- Terminated
- 2017-03-01
- Distribution
- Worldwide Distribution - US including MO, CA, OR, GA. NH, NY, MI, TX, MD, PA, IL, KS, NE, CT, AZ and Internationally to United Kingdom and Germany
- Recall number
- Z-0328-2017
- Recall status
- Terminated
- CDRH record ID
- 150382
- FDA classified
- 2016-11-09
- Recall event number
- 75395