- Recall classification
- Class II
Reason stated in the source
The Locking Punch Guide was not sufficiently guiding the large press fit punch, causing the implant to be placed 2-3mm more posterior than intended.
Source facts
- Action
- The field safety notice was disseminated by email on 12/12/2017. The notice stated the following: "DJO Sales Agents Action: Please review the bulletin that is provided with this notice. This bulletin contains new instructions on the use of the Locking Punch Guide and Punch. Please ensure your surgeons are familiar with this information. DJO Surgical then requires that you complete the attached acknowledgment and email it to teffany.hutto@djoglobal.com. Please have this form returned by December 15, 2017 Once redesigned guides are available, they will be exchanged through a separate field action."
- Initiated
- 2017-12-12
- Root cause
- Device Design
- Terminated
- 2022-06-13