Open original FDA record ↗Recall record
- FDA identifier
- Z-0328-2027
- Company in FDA record
- GE Medical Systems, LLC
- FDA classification date
- Nov 4, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).
Source facts
- Action
- GE HealthCare will send a customer letter dated August 28, 2026, that provides safety instructions for immediate mitigation of the incorrect configuration of the collimator anode track / filter in manual mode during 3D+2D acquisitions. GE HealthCare will correct all affected Senographe Pristina VIA systems by upgrading the software at no cost to the customer. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
- Initiated
- 2026-08-28
- Root cause
- Radiation Control for Health and Safety Act
- Distribution
- USA and OUS.
- Recall number
- Z-0328-2027
- Recall status
- Open, Classified
- CDRH record ID
- 223534
- FDA classified
- 2026-11-04
- Recall event number
- 100070